Aspects of gravimeter calibration
by time domain comparison of
gravity records
Bruno Meurers
Institute for Meteorology and Geophysics
University of Vienna
bruno.meurers@univie.ac.at
Abstract
This
paper addresses the accuracy problem of gravimeter calibration performed by
time domain comparison of different
gravity sensors. Un-modelled instrumental drift and the noise of the data are
main sources of systematic error components. Several synthetic and real case
studies are discussed to estimate accuracy limits. A
simple drift elimination method is proposed that is well suited to be applied
also for spring gravimeters with irregular drift. At least strong drift
components have to be removed otherwise
both calibration factor and time lag do not necessarily converge towards the
figures within the required 0.1% accuracy limit.
Introduction
High calibration
accuracy is still an important issue for getting reliable results in tidal
research, gravity monitoring and microgravimetry. Gravimeter calibration can be
done either in time or in frequency domain by comparing the instrumental response of two
sensors on common signals (e.g. earth tides, artificial gravitational or
inertial effects). The most important requirement is that signals of same
physical origin are compared only and that the sensor transfer functions are
considered. Time domain calibration
methods like regression analysis can be applied successfully even on short data
sets (< 200 h). Contrary, the frequency domain
calibration method requires long observation periods (³ 720 h) in order to
separate the main tidal constituents properly. Therefore
it detects calibration factor variations in much lower temporal resolution than
regression analysis, while drift determination, noise and different air
pressure response of the sensors are less critical.
A severe problem
is that the signal composition of both sensors differs due
to following reasons:
• instrumental noise and response on micro-seismic noise
• instrumental drift
• transfer function introducing
different time lags
• response on air pressure variations (e.g.
non-compensated Archimedian forces in LCR gravimeters)
Absolute gravimeters (AG) are commonly used as reference
sensors to calibrate superconducting
gravimeters (SG). Experience has shown that long data series of up to 7 days’
interval are necessary to get stable results with accuracy better than 0.1%
(e.g. Francis 1997, Francis et al. 1998). SGs exhibit an extremely small and
almost linear instrumental drift of less than a few µGal per year. Anti-alias
filters and 1 Hz sampling permit additional numerical filtering of the SG
output channel to obtain low noise data. Contrary, AG data is acquired with a much
longer sampling interval (15 – 30 s) and generally shows a much larger scatter.
Due to instrumental effects the existence of small drift components can be
excluded neither in SG nor in AG
records which could influence the calibration result systematically. Fig. 1 compares
the data of a calibration experiment performed in Vienna on 19990925. It demonstrates the
different noise level of the data sets used, but also clear systematic effects
in the AG data. This paper tries to address the influence of systematic effects
on the calibration result if they remain un-modelled prior to regression
analysis. This is done in a more general aspect in order to get accuracy limits
not only for AG-SG intercomparisons, but also for other
gravity sensor combinations including spring type gravimeters (e.g. LCR,
Scintrex). In this case the strong and irregular drift of spring gravimeters is
expected to introduce systematic
calibration errors. In addition, LCR gravimeters are known to give an abnormal
response on air pressure variations (e.g. Arnoso et al. 2001).

Fig.
1: Comparison of detided SG (bottom: GWR C025, 1s samples) and AG (top:
Jilag-6, 25s samples) data (Vienna,
19990925)
Synthetic studies
Several test calculations have been
performed to investigate the effect of
·
random noise and
time lag
- instrumental drift and
different air pressure response
using a data set based on predicted
model tides with 20s sampling. This sampling rate is typical for AG data. Both calibration factor and time
lag were determined by LSQ-adjustment.
Random
noise and time lag
The model
tides were compared with two different data sets:
- model
tides with time lag of 20s
- model
tides with time lag of 20s and normally distributed noise with a standard deviation
of 50 nms-2)
Several
different noise models have been used. As long as the standard deviation is not
larger than 50 nms-2, in each
case the adjusted calibration factor fulfils the 0.1% accuracy requirement.
However, convergence is very slow or even does not result exactly to the
expected figure. This highly depends on the noise structure. The same is valid
for the adjusted time lag. In addition, time lag adjustment does not
essentially improve the result of the calibration factor adjustment. Obviously
small un-modelled time lags do not influence the result strongly in spite of
the fact, that neglecting
different sensor time lags is equivalent to an additional signal consisting of diurnal,
semidiurnal and long-periodic components. If there
is no noise present in the data, the adjusted calibration factor is identical
almost exactly with the expected one even when the time lag is not adjusted.
Adjusted time lags correspond exactly to the expected ones. Fig. 2 is shown as
an example.

Fig. 2: Influence of phase shift and noise on the adjusted
calibration factor. Adjustment results are shown in dependence upon the number
of samples used. Both sensors’ data consist of model tides (20s samples); those
of the 2nd sensor have a time shift of 20s. Grey dots indicate the
results obtained when random noise
is superimposed to the data of the second sensor.
Instrumental drift and different
air pressure response
A major
problem is the presence of instrumental drift in the compared data sets,
because drift separation in the time domain
is a difficult task. Francis and Hendrickx (2001) applied a simultaneous
adjustment of the calibration factor and a third degree drift polynomial when
calibrating a LCR gravimeter by collocated SG observations. They achieved
temporarily stable accuracy of about 0.1% by analysing 15 days’ intervals.
However, the drift behaviour of some spring gravimeters does not permit low
degree polynomial adjustment. For those cases another
method is proposed here. It is based
on the approach by Lassovsky (1956) who used the zeros of model tides as
supporting points of the drift function. In this study a similar procedure has been applied. After subtracting the air
pressure effect by using a single admittance model, gravity readings at
moments when the model tides are zero yield the drift supporting points.
Finally a continuous drift function is constructed by cubic spline
interpolation.
In order to investigate both the efficiency of
this method and the effect of un-modelled drift components, several test
calculations have been performed by comparing model tides (20s
samples) to those with
different drift models superimposed. The drift models consist of both a linear
and a random component:
- systematic component: 5 nms-2/14
days,
random component: 10 supporting points/14 days,
standard deviation 5 nms-2
- systematic
component: 5 nms-2/14 days,
random component: 30 supporting points/14 days,
standard deviation 5 nms-2
- systematic
component: 30 nms-2/14 days,
random component: 10 supporting points/14 days,
standard deviation 5 nms-2
where the number of supporting points controls the
frequency content of the drift model. The selected drift parameter enables to
study the effect of even very irregular instrumental drift like that of LCR
gravimeters.
The examples shown in
Fig. 3 prove the drift elimination method to work properly even when high
frequency drift components are present. If the drift is eliminated before adjusting
the calibration factor, the latter converges very quickly towards the expected
figure. High frequency drift components make convergence worse. In this case a
time interval of about 6-8 days is required to get accurate results. If the
drift is not subtracted, the error of the adjusted calibration factor remains
below the 0.1% accuracy level after one week observation period except when
high systematic drift components are present.
The dependence of the
calibration result on the pre-processing method is tested finally by using real
gravity data from GWR C025. Model tides derived by tidal analysis of a 6.5
years’ recording of this SG in Vienna
served as reference signal. As second sensor, SG data sets covering a 14 days’ interval each
were applied after different kind of pre-processing:
1. no corrections
- air
pressure correction (single admittance model), but no drift elimination
- no
air pressure correction prior to drift elimination
- air
pressure correction prior to drift elimination
All data sets were decimated to 20s samples. Fig. 4
(top) shows the residuals after
subtracting the drift for the case studies 4 (black) and 3 (grey) respectively.
It proves that considering the air pressure effect is a necessary step to get
more reliable drift functions. If this effect is not corrected, it remains as
high frequency drift signal in the data and therefore
sometimes cannot be fully eliminated by the proposed method. This aspect is
important, if the two sensors are expected to respond on air pressure
differently (e.g. LCR gravimeters). The results of the calibration factor
adjustment are displayed in Fig. 4 (bottom). If no correction is performed at
all, both the calibration factor and time lag converge after an about 7 days’
observation interval, but to wrong figures. Fast and stable convergence occurs
only after removing the air pressure effect and instrumental drift. Reliable
time lags are obtained only if the air pressure effect is subtracted and if
drift remains untouched. Drift elimination corrupts the time lag information of
the data. As mentioned before, time shifted data can be composed of the
original one and of a systematic drift consisting of semidiurnal, diurnal and
long period components that are removed at least partially by drift
elimination.
The time domain calibration method is well suited to
determine calibration factor variations in high temporal resolution. This is
demonstrated by the last case study. During a more than 1-year period started
in June 2000, the LCR D-9 gravimeter equipped with a SRW-D type feedback
system (Schnüll et al. 1984) was monitoring parallel to the GWR C025 in Vienna. Tidal analyses of
successive, non-overlapping periods prove that the calibration factor of GWR
C025 is very constant (Meurers 2001). The amplitude factors for the main tidal
waves vary by less than 0.1% even when intervals as short as 1 month are
analysed (Fig. 6, open squares). Therefore
the SG can be used as stable reference to calibrate the feedback.

Fig. 3: Influence of un-modelled drift on the
adjusted calibration factor. Adjustment results are shown in dependence upon
the number of observation intervals used and the pre-processing method. Both
sensors’ data consist of model tides (20s samples); those of the 2nd
sensor are superimposed by instrumental drift. Different drift models have been
applied (see text).
pre-processing
method:
a)
drift correction
b)
no drift correction

Fig. 4: Influence of different
pre-processing steps on the calibration factor adjustment.
Bottom: Adjustment
results are shown in dependence upon the number of observations used and the
pre-processing method. 1st sensor: model tides derived by tidal
analysis of a 6.5 years’ recording of GWR C025. 2nd sensor: GWR C025
data (19970302 – 19970316), decimated to 20s samples.
Top: gravity residuals calculated by subtracting the drift. Air
pressure has been
(black) or has not been considered prior to drift
elimination (single admittance model).
The feedback calibration factor turned out to
be extremely unstable in time probably due
to a still unknown malfunction of its electronics. The LCR D-9 as a spring-type
gravimeter exhibits strong and irregular instrumental drift. In addition, its
response on air pressure variations differs significantly from that of the SG.
The admittance factor results to –5 nms-2/hPa instead of –3.5 nms-2/hPa.
Therefore the drift of both sensors
has been eliminated after air pressure correction applying the respective
admittance factors. Prior to this step both data sets were decimated to 5 min
samples. Successive overlapping intervals covering 2000 samples each
(approximately 7 days) have been analysed. Fig. 5 shows the temporal variations
of the feedback calibration factor resulting from the single adjustments.
The long-term behaviour of this variation can
be recognized also in Fig. 6 (grey dots), where
the amplitude factors of M2 and O1 are plotted versus
time. The latter were calculated by performing tidal analyses of successive
1-month intervals evaluated by using a constant feedback calibration factor.
Common features indicate sensitivity variations to be the reason. When taking
the temporal sensitivity variation according to Fig. 5 into account, the
amplitude factors get much more stable, especially in case of M2,
and common features disappear (Fig. 6, black dots).

Fig. 5: Calibration factor of LCR
D-9/SRW-D resulting from adjustments of successive intervals of 7 days (2000
samples, 5 min sampling) using GWR C025 data as reference.

Fig. 6: Amplitude factors resulting from
tidal analyses of successive intervals (1 month) recorded by LCR D‑9/SRW-D
and GWR C025.
Conclusions
The time domain
calibration method has limited accuracy. The main sources of systematic error
components are the noise of the data and un-modelled instrumental drift. For noisy
data the calibration factor converges very slowly with increasing number of
observations involved, but does not necessarily result exactly to the correct
figure, depending on the noise structure. However, the adjusted calibration
factor fulfils the 0.1% accuracy requirement. The same is valid for the time lag adjustment
that does not essentially improve the result of the calibration factor.
If strong drift
components are not removed both calibration factor and time lag do not converge
towards the correct figures even after observation periods longer than 7 days.
On the other hand, drift elimination
does no longer permit a time lag adjustment because it corrupts the phase
information of the data.
If the compared sensors exhibit low and regular
drift like SGs, accuracy better than 0.1% can be obtained from data covering an
interval of 6-8 days. Although AG data sometimes show a small apparent drift
caused by time dependent systematic effects, drift elimination is not
recommended when calibrating a SG by comparing with AG data, as it removes
physical signal components (e.g. air pressure effect) at least partially and
perhaps differently for both instruments.
The situation is quite different when
calibrating a spring-type gravimeter by comparison with SG data. Spring
gravimeters often show strong and irregular instrumental drift and different
response to air pressure variations. In this case the drift has to be
eliminated before the regression analysis, and the air pressure effect has to
be subtracted for both sensors before drift determination.
References
Arnoso, J.,
Vieira, R., Velez, E.J., Van Ruymbeke, M. and Venedikov, A.P., 2001: Studies of
tides and instrumental performance of three gravimeters at Cueva de los Verdes
(Lanzarote, Spain).
Jour. Geod. Soc. Japan,
47, 1, 70-75.
Francis,
O., 1997: Calibration of the C021 Superconducting
Gravimeter in Membach (Belgium)
using 47 days of absolute gravity measurements. IAG Symposia, 117, 212-219.
Francis,
O., Niebauer, T.M., Sasagawa, G., Klopping, F. and Gschwind, J., 1998:
Calibration of a superconducting
gravimeter by comparison with an absolute gravimeter FG5 in Boulder, Geophys. Res. Lett., 25, 1075-1078.
Francis,
O. and Hendrickx, M., 2001: Calibration of the LaCoste-Romberg 906 by Comparison
with the Superconducting Gravimeter
C021 in Membach (Belgium),
Jour. Geod. Soc. Japan, 47,
1, 16-21.
Lassovsky,
K., 1956: A Fold deformacios egyutthatojanak meghatarozasa
gravimeteresz-lelesekbol. Geofisikai Kozlemenyek, 2, 18-26.
Meurers, B., 2001: Tidal and
Non-tidal Gravity Variations in Vienna
- a Five Years‘ SG Record, Jour. Geod. Soc. Japan, 47, 1, 392-397.
Schnüll, M., Röder,
R.H. and Wenzel, H.G., 1984: An improved electronic feedback system for
LaCoste&Romberg gravity meters. BGI, Bulletin d'Information, 55, 27-36.